ENZYME-IMMUNOASSAY FOR MYOGLOBIN IN HUMAN-SERUM AND URINE - METHOD DEVELOPMENT, NORMAL VALUES AND APPLICATION TO ACUTE MYOCARDIAL-INFARCTION

被引:20
作者
CLOONAN, MJ
BISHOP, GA
WILTONSMITH, PD
CARTER, IW
ALLAN, RM
WILCKEN, DEL
机构
[1] PRINCE HENRY HOSP,DIV MICROBIOL,SYDNEY,NEW S WALES,AUSTRALIA
[2] PRINCE HENRY HOSP,DIV CARDIOVASC MED,SYDNEY,NEW S WALES,AUSTRALIA
关键词
D O I
10.3109/00313027909059049
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
An enzyme-immunoassay has been developed for the detection of myoglobin in human serum and urine which is specific, accurate, precise, and has a sensitivity of 3 ng/ml. When compared with radioimmunoassay, the enzyme-immunoassay gives markedly similar results. Sera from normal adults had a myoglobin concentration in the range 3-65 ng/ml, and 64% of the same group had detectable myoglobinuria (range 3-11.5 ng/ml). All of 8 patients with definite acute myocardial infarction had raised serum myoglobin levels (range 200-1125 ng/ml) either at admission or 4 h later. Myoglobin concentrationreturned to normal in 6 patients, and in the remaining 2 patients there was evidence of infarct extension. Urinary myoglobin excretion was variable. One patient with possible acute myocardial infarction had elevated serum myoglobin (413 ng/ml 4 h post admission) and 5 patients with no evidence of infarction had normal levels (15-53 ng/ml). The results suggest that detection of serum myoglobin by enzyme-immunoassay may be a valuable test in the early diagnosis of acute myocardial infarction. © 1979 Royal College of Pathologists of Australasia.
引用
收藏
页码:689 / 699
页数:11
相关论文
共 40 条
[21]  
MIYOSHI K, 1978, J LAB CLIN MED, V92, P341
[22]  
NEVINS, 1973, JAMA-J AM MED ASSOC, V224, P1382
[23]   RADIOIMMUNOASSAY FOR HUMAN MYOGLOBIN - INITIAL EXPERIENCE IN PATIENTS WITH CORONARY HEART-DISEASE [J].
REICHLIN, M ;
VISCO, JP ;
KLOCKE, FJ .
CIRCULATION, 1978, 57 (01) :52-56
[24]   MYOCARDIAL-INFARCT EXTENSION DETECTED BY PRECORDIAL ST-SEGMENT MAPPING [J].
REID, PR ;
TAYLOR, DR ;
KELLY, DT ;
WEISFELDT, ML ;
HUMPHRIES, JO ;
ROSS, RS ;
PITT, B .
NEW ENGLAND JOURNAL OF MEDICINE, 1974, 290 (03) :123-128
[25]   STATISTICAL QUALITY CONTROL OF RADIOIMMUNOASSAYS [J].
RODBARD, D ;
RAYFORD, PL ;
COOPER, JA ;
ROSS, GT .
JOURNAL OF CLINICAL ENDOCRINOLOGY & METABOLISM, 1968, 28 (10) :1412-+
[26]  
ROE CR, 1977, CLIN CHEM, V23, P1807
[27]  
ROSANO TG, 1977, CLIN CHEM, V23, P868
[28]  
ROSANO TG, 1977, CLIN CHEM, V23, P69
[29]   NEW DIAGNOSTIC TEST FOR ACUTE MYOCARDIAL-INFARCTION - DETECTION OF MYOGLOBINURIA BY RADIOIMMUNODIFFUSION ASSAY [J].
SARANCHA.HJ ;
BERNSTEI.SH .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1974, 228 (10) :1251-1255
[30]   ENZYME-IMMUNOASSAY [J].
SCHUURS, AHWM ;
VANWEEMEN, BK .
CLINICA CHIMICA ACTA, 1977, 81 (01) :1-40