FIELD-CORRECTED ION CELL FOR ION-CYCLOTRON RESONANCE

被引:66
作者
HANSON, CD
CASTRO, ME
KERLEY, EL
RUSSELL, DH
机构
[1] TEXAS A&M UNIV SYST,DEPT CHEM,COLLEGE STN,TX 77843
[2] UNIV NO IOWA,DEPT CHEM,CEDAR FALLS,IA 50614
关键词
D O I
10.1021/ac00204a018
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The basis for mass analysis by Fourier transform ion cyclotron resonance (FT-ICR) is the motion of ions in a homogeneous magnetic field. Electric fields are used In FT-ICR to modify ion motion (i.e., trapping and excitation fields). Ion interaction with inhomogeneous radio frequency and direct current (dc) electric fields complicates the ion motion and leads to loss of both resolution and sensitivity Modifying the geometry of the FT-ICR cell to produce uniform ion acceleration and homogeneous dc trapping fields resuits in simplified frequency measurements and studies of ion-molecule reactions. © 1990, American Chemical Society. All rights reserved.