TRANSVENOUS-SUBCUTANEOUS DEFIBRILLATION LEADS - EFFECT OF TRANSVENOUS ELECTRODE POLARITY ON DEFIBRILLATION THRESHOLD

被引:11
作者
BLOCK, M
HAMMEL, D
BOCKER, D
BORGGREFE, M
BUDDE, T
ISBRUCH, F
SCHELD, HH
BREITHARDT, G
机构
[1] UNIV MUNSTER HOSP,DEPT CARDIOL ANGIOL,MUNSTER,GERMANY
[2] DEPT CARDIOVASC SURG,MUNSTER,GERMANY
[3] INST RES ARTERIOSCLEROSIS,MUNSTER,GERMANY
关键词
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR; VENTRICULAR FIBRILLATION; ENDOCARDIAL DEFIBRILLATION; DEFIBRILLATION THRESHOLD;
D O I
10.1111/j.1540-8167.1994.tb01131.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: The defibrillation threshold (DFT) of a transvenous-subcutaneous electrode configuration is sometimes unacceptably high. To obtain a DFT with a sufficient safety margin, the defibrillation held can be modified by repositioning the electrodes or more easily by a change of electrode polarity. In a prospective randomized cross-over study, the effect of transvenous electrode polarity on DFT was evaluated. Methods and Results: In 21 patients receiving transvenous-subcutaneous defibrillation leads, the DFT was determined intraoperatively for two electrode configurations. Two monophasic defibrillation pulses were delivered in sequential mode between either the right ventricular (RV) electrode as common cathode and the superior vena cava (SVC) and subcutaneous electrodes as anodes (configuration I) or the SVC electrode as common cathode and the RV and subcutaneous electrodes as anodes (configuration II). In each patient, both electrode configurations were used alternately with declining energies (25, 15, 10, 5, 2 J) until failure of defibrillation occurred. The DFT did not differ between both configurations (18.3 +/- 8.2 J vs 18.9 +/- 8.9 J; P = 0.72). Eleven patients had the same DFT with both electrode configurations, 5 patients a lower DFT with the RV electrode as cathode, and 5 patients a lower DFT with the SVC as cathode. Four patients had a sufficiently low DFT (less than or equal to 25 J) with only 1 of the 2 configurations. Conclusion: A change of electrode polarity of transvenous-subcutaneous defibrillation electrodes may result in effective defibrillation if the first electrode polarity tested fails to defibrillate. In general, neither the RV electrode nor the SVC electrode is superior if used as a common cathode in combination with a subcutaneous anodal chest patch.
引用
收藏
页码:912 / 918
页数:7
相关论文
共 26 条
[21]  
Pachulski Roman T., 1993, Journal of the American College of Cardiology, V21, p126A
[22]   LOW-ENERGY ENDOCARDIAL DEFIBRILLATION USING AN AXILLARY OR A PECTORAL THORACIC ELECTRODE LOCATION [J].
SAKSENA, S ;
DEGROOT, P ;
KROL, RB ;
RAJU, R ;
MATHEW, P ;
MEHRA, R .
CIRCULATION, 1993, 88 (06) :2655-2660
[23]   SIMULTANEOUS BIPHASIC SHOCKS ENHANCE EFFICACY OF ENDOCARDIAL CARDIOVERSION DEFIBRILLATION IN MAN [J].
SAKSENA, S ;
AN, H ;
KROL, RB ;
BURKHARDT, E .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (11) :1935-1942
[24]   ENDOCARDIAL PACING, CARDIOVERSION AND DEFIBRILLATION USING A BRAIDED ENDOCARDIAL LEAD SYSTEM [J].
SAKSENA, S ;
LUCERI, R ;
KROL, RB ;
BROWNSTEIN, S ;
BURKHARDT, E ;
ACCORTI, P ;
BREWER, G ;
SCOTT, S ;
CALLAGHAN, F ;
LIVINGSTON, A .
AMERICAN JOURNAL OF CARDIOLOGY, 1993, 71 (10) :834-841
[25]  
STRICKBERGER SA, 1994, J AM COLL CARDIOL, V23, pA12
[26]   COMPARISON OF BIPHASIC AND MONOPHASIC SHOCKS FOR DEFIBRILLATION USING A NONTHORACOTOMY SYSTEM [J].
WYSE, DG ;
KAVANAGH, KM ;
GILLIS, AM ;
MITCHELL, LB ;
DUFF, HJ ;
SHELDON, RS ;
KIESER, TM ;
MAITLAND, A ;
FLANAGAN, P ;
ROTHSCHILD, J ;
MEHRA, R .
AMERICAN JOURNAL OF CARDIOLOGY, 1993, 71 (02) :197-202