Accuracy of an electromagnetic measurement device and application to the measurement and description of knee joint motion

被引:56
作者
Bull, AMJ [1 ]
Berkshire, FH [1 ]
Amis, AA [1 ]
机构
[1] Univ London Imperial Coll Sci Technol & Med, Dept Mech Engn, Biomech Sect, London SW7 2AZ, England
关键词
electromagnetic measurement device; knee joint motion; clinical rotations;
D O I
10.1243/0954411981534123
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The aims of this study were to analyse the accuracy and sensitivity to metals of an electromagnetic motion measurement device and to use it to provide a description of motion of the human knee joint. A calibrated grid and sine bar were used to measure translational and rotational accuracy, which were found to be +/-0.23 per cent of the step size (translation) and +/-2.0 per cent of the step size (rotation) within an optimal operational zone of minimal positional error for which the receiver to transmitter separation was between 271 and 723 mm. The presence of multiple receivers was found to have a significant effect on the accuracy only when positioned within 30 mm of one another. Mild steel was found to have a significant detrimental effect when within 150 mm of the transmitter or receiver. A stainless steel bone screw had no effect on the accuracy of the device. A mathematical description of knee joint motion is presented which describes the motion in terms of clinical rotations. The device is a useful tool for measuring joint motion in the operating theatre and laboratory, owing to its accuracy and insensitivity to surgical alloys.
引用
收藏
页码:347 / 355
页数:9
相关论文
共 23 条
[1]   INVITRO LIGAMENT TENSION PATTERN IN THE FLEXED KNEE IN PASSIVE LOADING [J].
AHMED, AM ;
HYDER, A ;
BURKE, DL ;
CHAN, KH .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1987, 5 (02) :217-230
[2]   APPLICATION OF A MAGNETIC TRACKING DEVICE TO KINESIOLOGIC STUDIES [J].
AN, KN ;
JACOBSEN, MC ;
BERGLUND, LJ ;
CHAO, EYS .
JOURNAL OF BIOMECHANICS, 1988, 21 (07) :613-&
[3]   3-DIMENSIONAL KINEMATICS OF GLENOHUMERAL ELEVATION [J].
AN, KN ;
BROWNE, AO ;
KORINEK, S ;
TANAKA, S ;
MORREY, BF .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1991, 9 (01) :143-149
[4]  
BULL AMJ, 1996, T EUR SOC BIOMECH, V10, P368
[5]  
BULL AMJ, 1997, T ORTHOP RES SOC, V22, P873
[6]   THE USE OF ROBOTICS TECHNOLOGY TO STUDY HUMAN JOINT KINEMATICS - A NEW METHODOLOGY [J].
FUJIE, H ;
MABUCHI, K ;
WOO, SLY ;
LIVESAY, GA ;
ARAI, S ;
TSUKAMOTO, Y .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1993, 115 (03) :211-217
[7]  
Goldstein H, 1980, CLASSICAL MECH
[8]   A JOINT COORDINATE SYSTEM FOR THE CLINICAL DESCRIPTION OF 3-DIMENSIONAL MOTIONS - APPLICATION TO THE KNEE [J].
GROOD, ES ;
SUNTAY, WJ .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1983, 105 (02) :136-144
[9]   MEASUREMENT OF TOTAL MOTION BETWEEN 2 BODY SEGMENTS .1. ANALYTICAL DEVELOPMENT [J].
KINZEL, GL ;
HILLBERRY, BM ;
HALL, AS .
JOURNAL OF BIOMECHANICS, 1972, 5 (01) :93-+
[10]   6R INSTRUMENTED SPATIAL LINKAGES FOR ANATOMICAL JOINT MOTION MEASUREMENT .1. DESIGN [J].
KIRSTUKAS, SJ ;
LEWIS, JL ;
ERDMAN, AG .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1992, 114 (01) :92-100