Simultaneous determination of p-aminohippuric acid, acetyl-p-aminohippuric acid and iothalamate in human plasma and urine by high-performance liquid chromatography

被引:25
作者
Dowling, TC [1 ]
Frye, RF [1 ]
Zemaitis, MA [1 ]
机构
[1] Univ Pittsburgh, Sch Pharm, Dept Pharmaceut Sci, Pittsburgh, PA 15261 USA
来源
JOURNAL OF CHROMATOGRAPHY B | 1998年 / 716卷 / 1-2期
关键词
p-aminohippuric acid; acetyl-p-aminohippuric acid; iothalamate;
D O I
10.1016/S0378-4347(98)00294-1
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A sensitive and specific high-performance liquid chromatographic assay was developed for the simultaneous determination of p-aminohippuric acid (PAB), acetyl-p-aminohippuric acid (aPAH), and iothalamate in human plasma and urine. Plasma samples were prepared by protein precipitation with acetonitrile followed by evaporation, reconstitution in mobile phase, and injection onto a C-18 reversed-phase column. Urine samples were diluted with 3 volumes of mobile phase prior to injection. Column effluent was monitored by UV detection at 254 nm. The lower limits of quantification in plasma were 0.5 mg/l for PAH and aPAH, and 1.0 mg/l for iothalamate. The within-day and between-day coefficients of variation in plasma and urine were less than or equal to 7.8% for all analytes. This method is well suited for renal function studies using iothalamate and PAH, whether administered as a bolus dose or by continuous infusion, to measure glomerular filtration rate and effective renal plasma flow, respectively. (C) 1998 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:305 / 313
页数:9
相关论文
共 26 条
[1]  
Agarwal R, 1998, J CHROMATOGR B, V705, P3
[2]  
AGODOA L, 1997, KIDNEY INT S63, V52, pS144
[3]  
Apperloo AJ, 1996, J AM SOC NEPHROL, V7, P567
[4]   Effect of black tea drinking on blood lipids, blood pressure and aspects of bowel habit [J].
Bingham, SA ;
Vorster, H ;
Jerling, JC ;
Magee, E ;
Mulligan, A ;
Runswick, SA ;
Cummings, JH .
BRITISH JOURNAL OF NUTRITION, 1997, 78 (01) :41-55
[5]   HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD FOR THE SIMULTANEOUS DETERMINATION OF IOTHALAMATE AND OMICRON-IODOHIPPURATE [J].
BOSCHI, S ;
MARCHESINI, B .
JOURNAL OF CHROMATOGRAPHY, 1981, 224 (01) :139-143
[6]   Specific determination of PAH and its N-acetyl metabolite by HPLC increases the accuracy and precision of PAH clearance measurements [J].
Decosterd, LA ;
Bélaz, N ;
Appenzeller, M ;
Maghraoui, A ;
Rousso, P ;
Buclin, T ;
Biollaz, J .
RENAL FAILURE, 1998, 20 (02) :311-317
[7]   High-performance liquid chromatography of the renal blood flow marker p-aminohippuric acid (PAH) and its metabolite N-acetyl PAH improves PAH clearance measurements [J].
Decosterd, LA ;
Karagiannis, A ;
Roulet, JM ;
Bélaz, N ;
Appenzeller, M ;
Buclin, T ;
Vogel, P ;
Biollaz, J .
JOURNAL OF CHROMATOGRAPHY B, 1997, 703 (1-2) :25-36
[8]  
Dowling T. C., 1997, Clinical Pharmacology and Therapeutics, V61, P215
[9]  
FRENNBY B, 1995, CLIN NEPHROL, V43, P35
[10]  
GASPARI F, 1997, KIDNEY INT S, V63, pS151