THE CONTINUITY EQUATION TESTED IN A BILEAFLET AORTIC PROSTHESIS

被引:21
作者
CHAMBERS, J [1 ]
COPPACK, F [1 ]
DEVERALL, P [1 ]
JACKSON, G [1 ]
SOWTON, E [1 ]
机构
[1] GUYS HOSP,DEPT CARDIOTHORAC SURG,LONDON SE1 9RT,ENGLAND
关键词
CONTINUITY EQUATION; DOPPLER ULTRASOUND; PROSTHETIC VALVE;
D O I
10.1016/0167-5273(91)90209-8
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The continuity equation is valid for a population of aortic valves, but its accuracy in individual valves is uncertain. In bileaflet prostheses, obstruction to forward flow is small and individual variability in opening behaviour is minimal. In these valves, the area of the effective orifice should be close to that measured by the manufacturer. A total of 57 patients aged 58 +/- 11 years were studied at a mean of 3.6 months after implantation with a CarboMedics aortic prosthesis. Nine had additional implants in the mitral position and all prostheses were clinically normal. Peak subaortic and transaortic velocities were averaged over 5 beats. The area of the effective orifice was significantly different between the four diameters (P < 0.00001), and the correlation between the effective and measured area of the orifice was moderate (r(s) = 0.73, P < 0.00003). The 95% range for the differences between individual pairs of values, however, was 0.16 +/- 0.61 cm2. Discrepancies probably arose in the estimation of subaortic cross-sectional area and subaortic velocity. Thus, the continuity equation may be inaccurate in an individual prosthetic valve when functioning normally.
引用
收藏
页码:149 / 154
页数:6
相关论文
共 15 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]  
BUTTERFIELD M, 1990, 7TH P M EUR SOC BIOM
[3]  
CHAMBERS JB, UNPUB J AM COLL CARD
[4]   DOPPLER ULTRASOUND OF NORMALLY FUNCTIONING MECHANICAL MITRAL AND AORTIC-VALVE PROSTHESES [J].
GIBBS, JL ;
WHARTON, GA ;
WILLIAMS, GJ .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 1988, 18 (03) :391-398
[5]   NONINVASIVE EVALUATION OF AORTIC-STENOSIS SEVERITY UTILIZING DOPPLER ULTRASOUND AND ELECTRICAL BIOIMPEDANCE [J].
GOLI, VD ;
TEAGUE, SM ;
PRASAD, R ;
HARVEY, J ;
VOYLES, WF ;
OLSON, EG ;
SCHECHTER, E ;
THADANI, U .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1988, 11 (01) :66-71
[7]   HYDRAULIC FORMULA FOR CALCULATION OF THE AREA OF THE STENOTIC MITRAL VALVE, OTHER CARDIAC VALVES, AND CENTRAL CIRCULATORY SHUNTS .1. [J].
GORLIN, R ;
GORLIN, SG .
AMERICAN HEART JOURNAL, 1951, 41 (01) :1-29
[8]   PULSED DOPPLER ECHOCARDIOGRAPHIC DETERMINATION OF STROKE VOLUME AND CARDIAC-OUTPUT - CLINICAL VALIDATION OF 2 NEW METHODS USING THE APICAL WINDOW [J].
LEWIS, JF ;
KUO, LC ;
NELSON, JG ;
LIMACHER, MC ;
QUINONES, MA .
CIRCULATION, 1984, 70 (03) :425-431
[9]   REPRODUCIBILITY OF ECHOCARDIOGRAPHIC ESTIMATES OF THE AREA OF STENOSED AORTIC VALVES USING THE CONTINUITY EQUATION [J].
MYRENG, Y ;
MOLSTAD, P ;
ENDRESEN, K ;
IHLEN, H .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 1990, 26 (03) :349-354
[10]   PREDICTION OF THE SEVERITY OF AORTIC-STENOSIS BY DOPPLER AORTIC-VALVE AREA DETERMINATION - PROSPECTIVE DOPPLER-CATHETERIZATION CORRELATION IN 100 PATIENTS [J].
OH, JK ;
TALIERCIO, CP ;
HOLMES, DR ;
REEDER, GS ;
BAILEY, KR ;
SEWARD, JB ;
TAJIK, AJ .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1988, 11 (06) :1227-1234