PHASE-VELOCITY MAPPING WITH A REAL-TIME LINE SCAN TECHNIQUE

被引:15
作者
BUTTS, K [1 ]
HANGIANDREOU, NJ [1 ]
RIEDERER, SJ [1 ]
机构
[1] MAYO CLIN & MAYO FDN,MAGNET RESONANCE LAB,200 1ST ST SW,ROCHESTER,MN 55905
关键词
LINE SCAN; PHASE DIFFERENCE; REAL-TIME FLOW MEASUREMENT;
D O I
10.1002/mrm.1910290125
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A real-time, 20-Hz, one-dimensional MR velocity imaging technique is described. A two-dimensional RF pulse excites a 3-cm diameter column. Velocity maps are formed from the phase difference between successive flow encoded and compensated acquisitions. A three-point subtraction variation provides reduced sensitivity to static spins.
引用
收藏
页码:134 / 138
页数:5
相关论文
共 26 条
[1]  
BERNSTEIN MA, 1991, J MAGN RESON IMAGING, V1, P6
[2]   FAST MR CARDIAC PROFILING WITH 2-DIMENSIONAL SELECTIVE PULSES [J].
CLINE, HE ;
HARDY, CJ ;
PEARLMAN, JD .
MAGNETIC RESONANCE IN MEDICINE, 1991, 17 (02) :390-401
[3]   QUANTITATIVE MEASUREMENT OF BLOOD-FLOW USING CYLINDRICALLY LOCALIZED FOURIER VELOCITY ENCODING [J].
DUMOULIN, CL ;
SOUZA, SP ;
HARDY, CJ ;
ASH, SA .
MAGNETIC RESONANCE IN MEDICINE, 1991, 21 (02) :242-250
[4]   SPATIAL PRESATURATION - A METHOD FOR SUPPRESSING FLOW ARTIFACTS AND IMPROVING DEPICTION OF VASCULAR ANATOMY IN MR IMAGING [J].
FELMLEE, JP ;
EHMAN, RL .
RADIOLOGY, 1987, 164 (02) :559-564
[5]   ECHO-PLANAR HIGH-RESOLUTION FLOW VELOCITY MAPPING [J].
FIRMIN, DN ;
KLIPSTEIN, RH ;
HOUNSFIELD, GL ;
PALEY, MP ;
LONGMORE, DB .
MAGNETIC RESONANCE IN MEDICINE, 1989, 12 (03) :316-327
[6]  
FRAYNE R, 1992, BOOK ABSTR SMRM, P2907
[7]  
GAO J, 1991, MED PHYS, V18, P3
[8]  
HANGIANDREOU NJ, IN PRESS J MAGN RESO
[9]  
HANGIANDREOU NJ, 1992, 10TH ANN SMRI, P142
[10]   RAPID NMR CARDIOGRAPHY WITH A HALF-ECHO M-MODE METHOD [J].
HARDY, CJ ;
PEARLMAN, JD ;
MOORE, JR ;
ROEMER, PB ;
CLINE, HE .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1991, 15 (05) :868-874