EVALUATION OF THE E-TEST, A NOVEL METHOD OF QUANTIFYING ANTIMICROBIAL ACTIVITY

被引:160
作者
BROWN, DFJ
BROWN, L
机构
[1] Clinical Microbiology and Public Health Laboratory, Addenbrooke's Hospital, Cambridge CB2 2QW, Hills Road
关键词
D O I
10.1093/jac/27.2.185
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
The 'E test' is a method for measuring MICs of antimicrobial agents against bacteria and is based on diffusion of a pre-formed antibiotic gradient from a plastic strip. The performance of the E test was evaluated by comparison with a conventional agar dilution MIC method in tests with ten agents and a variety of organisms. Correlation between MICs by the agar dilution and E test methods was good, 98.85% results were within 2 log2 dilution steps in a total of 1304 tests. The E test is technically straightforward as tests are set up in the same way as the disc diffusion method. The versatility and ease of use of the E test make the method an attractive alternative to conventional dilution tests. © 1991 The British Society for Antimicrobial Chemotherapy.
引用
收藏
页码:185 / 190
页数:6
相关论文
共 4 条
[1]  
BOLMSTROM A, 1988, 28TH INT C ANT AG CH, P325
[2]  
ERICSSON HM, 1971, ACTA PATHOLOG MICROB, P1
[3]  
Lurian V, 1986, ANTIBIOTICS LAB MED, P93
[4]   NOVEL METHOD FOR DETECTION OF BETA-LACTAMASES BY USING A CHROMOGENIC CEPHALOSPORIN SUBSTRATE [J].
OCALLAGHAN, CH ;
SHINGLER, AH ;
KIRBY, SM ;
MORRIS, A .
ANTIMICROBIAL AGENTS AND CHEMOTHERAPY, 1972, 1 (04) :283-+