LIMITATIONS OF THE ZERO CROSSING DETECTOR IN THE ANALYSIS OF INTRACORONARY DOPPLER - A COMPARISON WITH FAST FOURIER-TRANSFORM ANALYSIS OF BASAL, HYPEREMIC, AND TRANSSTENOTIC BLOOD-FLOW VELOCITY-MEASUREMENTS IN PATIENTS WITH CORONARY-ARTERY DISEASE

被引:10
作者
DIMARIO, C [1 ]
ROELANDT, JRTC [1 ]
DEJAEGERE, P [1 ]
LINKER, DT [1 ]
OOMEN, J [1 ]
SERRUYS, PW [1 ]
机构
[1] ERASMUS UNIV, THORAXCTR,DIV CARDIOL,CARDIAC CATHETERIZAT LAB, POB 1738, 3000 DR ROTTERDAM, NETHERLANDS
来源
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS | 1993年 / 28卷 / 01期
关键词
BLOOD FLOW VELOCITY; DOPPLER ULTRASOUND; CORONARY STENOSIS; CORONARY FLOW RESERVE;
D O I
10.1002/ccd.1810280112
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The current clinical standard for the analysis of intracoronary Doppler signals is the application of a zero-crossing (ZC) detector. However, the accuracy of the method is questionable, especially in areas of disturbed flow, as confirmed by in vitro studies, animal experiments, and intraoperative observations. The alm of this study is the comparison of a conventional ZC detector and a custom-designed spectral analyzer (fast Fourier transform, FFT) in the analysis of intracoronary Doppler signals obtained in 19 patients undergoing coronary angioplasty. A 3F catheter with an end-mounted Doppler ceramic crystal was placed over an 0.01 4'' guidewire in a normal or near-normal segment proximal to the lesion to be dilated. The Doppler signal was recorded before and after intracoronary infusion of 12.5 mg of papaverine. In 9 patients high flow velocities could be recorded when the catheter was advanced across the stenosis. The blood flow velocity measurements obtained with ZC were significantly lower than the maximal FFT flow velocity measurements (16 +/- 12 cm/s vs. 29 +/- 18 cm/s, p < .001). In all the conditions of Doppler signal acquisition (baseline, hyperemia, stenosis) a large scattering of the signed differences between corresponding measurements was observed. The standard deviation of the difference ZC-FFT was +/- 11 cm/s and +/- 5 cm/s for the maximal and mean FFT flow velocity, corresponding in both cases to +/- 37% of the mean of the ZC and FFT measurements. Large differences were also observed in the values of coronary flow reserve (CFR) calculated as the ratio between ZC and FFT flow velocity measurements 30 s after papaverine intracoronary and at baseline. The standard deviation of the difference ZC-FFT based CFR was +/- 1.3 and +/- 1.2 for the values derived from the maximal and mean FFT flow velocities (percent difference +/- 32% and +/- 37%, respectively). In conclusion, the measurements obtained from the same intracoronary Doppler signal analyzed with a ZC detector and an FFT technique showed large differences in various conditions of flow and also in the assessment of relative flow velocity derived indices such as CFR. Spectral analysis should replace the current use of a ZC detector for the evaluation of coronary Doppler signals, even for the assessment of relative flow velocity changes.
引用
收藏
页码:56 / 64
页数:9
相关论文
共 20 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   CYCLIC CHANGES OF BLOOD ECHOGENICITY IN HIGH-FREQUENCY ULTRASOUND [J].
DEKROON, MGM ;
SLAGER, CJ ;
GUSSENHOVEN, WJ ;
SERRUYS, PW ;
ROELANDT, JRTC ;
BOM, N .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1991, 17 (07) :723-728
[3]   THE RELATIONSHIP BETWEEN TIME AVERAGED INTENSITY WEIGHTED MEAN VELOCITY, AND TIME AVERAGED MAXIMUM VELOCITY IN NEONATAL CEREBRAL-ARTERIES [J].
EVANS, DH ;
SCHLINDWEIN, FS ;
LEVENE, MI .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1989, 15 (05) :429-435
[4]   DIVERGENT EFFECTS OF SEROTONIN ON CORONARY-ARTERY DIMENSIONS AND BLOOD-FLOW IN PATIENTS WITH CORONARY ATHEROSCLEROSIS AND CONTROL PATIENTS [J].
GOLINO, P ;
PISCIONE, F ;
WILLERSON, JT ;
CAPPELLIBIGAZZI, M ;
FOCACCIO, A ;
VILLARI, B ;
INDOLFI, C ;
RUSSOLILLO, E ;
CONDORELLI, M ;
CHIARIELLO, M .
NEW ENGLAND JOURNAL OF MEDICINE, 1991, 324 (10) :641-648
[5]  
HATLE L, 1982, DOPPLER ULTRASOUND C, P32
[6]   ASSESSMENT OF SEVERITY OF CORONARY STENOSES USING A DOPPLER CATHETER - VALIDATION OF A METHOD BASED ON THE CONTINUITY EQUATION [J].
JOHNSON, EL ;
YOCK, PG ;
HARGRAVE, VK ;
SREBRO, JP ;
MANUBENS, SM ;
SEITZ, W ;
PORTS, TA .
CIRCULATION, 1989, 80 (03) :625-635
[7]  
KAJIYA F, 1986, CIRCULATION, V74, P53
[8]  
Lunt M J, 1975, Ultrasound Med Biol, V2, P1, DOI 10.1016/0301-5629(75)90035-6
[9]   CORONARY BLOOD-FLOW VELOCITY DURING PERCUTANEOUS TRANS-LUMINAL CORONARY ANGIOPLASTY AS A GUIDE FOR ASSESSMENT OF THE FUNCTIONAL RESULT [J].
SERRUYS, PW ;
JUILLIERE, Y ;
ZIJLSTRA, F ;
BEATT, KJ ;
DEFEYTER, PJ ;
SURYAPRANATA, H ;
VANDENBRAND, M ;
ROELANDT, J .
AMERICAN JOURNAL OF CARDIOLOGY, 1988, 61 (04) :253-259
[10]  
SERRUYS PW, 1990, TXB INTERVENTIONAL C, P768