LIMITATIONS OF ULTRASONOGRAPHY IN SURVEILLANCE OF SMALL ABDOMINAL AORTIC-ANEURYSMS

被引:113
作者
ELLIS, M [1 ]
POWELL, JT [1 ]
GREENHALGH, RM [1 ]
机构
[1] CHARING CROSS & WESTMINSTER MED SCH,DEPT SURG,FULHAM PALACE RD,LONDON W6 8RF,ENGLAND
关键词
D O I
10.1002/bjs.1800780529
中图分类号
R61 [外科手术学];
学科分类号
摘要
The repeatability, observer bias and instrument bias of aortic diameter measurement by ultrasonography, were investigated in ten patients with small (3-6 cm by computed tomography) infrarenal abdominal aortic aneurysm. The repeatability of maximum aortic diameter measurement by ultrasonography was much better for anterior-posterior than transverse diameter, with coefficients of repeatability 3.0-7.5 mm and 10-15 mm respectively. The repeatability of suprarenal aortic diameter measurement was poor. Surprisingly, maximum diameter using ultrasonography was larger than that using computed tomography, the difference being least for anterior-posterior measurements. At best a single, experienced observer, using the same instrument may provide aortic diameters using ultrasonography accurate to within 5 mm, but more commonly such aortic diameter is only accurate to within 8 mm.
引用
收藏
页码:614 / 616
页数:3
相关论文
共 7 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]  
COLLIN J, 1988, LANCET, V2, P613
[3]  
LITTOOY FN, 1989, ARCH SURG-CHICAGO, V124, P419
[4]   PROGNOSIS OF ABDOMINAL AORTIC-ANEURYSMS - A POPULATION-BASED STUDY [J].
NEVITT, MP ;
BALLARD, DJ ;
HALLETT, JW .
NEW ENGLAND JOURNAL OF MEDICINE, 1989, 321 (15) :1009-1014
[5]   GENERAL PRACTICE-BASED POPULATION SCREENING FOR ABDOMINAL AORTIC-ANEURYSMS - A PILOT-STUDY [J].
OKELLY, TJ ;
HEATHER, BP .
BRITISH JOURNAL OF SURGERY, 1989, 76 (05) :479-480
[6]   ROUTINE ULTRASOUND SCREENING IN MANAGEMENT OF ABDOMINAL AORTIC-ANEURYSM [J].
SCOTT, RAP ;
ASHTON, HA ;
KAY, DN .
BRITISH MEDICAL JOURNAL, 1988, 296 (6638) :1709-1710
[7]  
1979, BS5497 BSI BRIT ST 1