CORRECTING SPATIAL DISTORTIONS AND NONUNIFORM RESPONSE IN AREA DETECTORS

被引:32
作者
STANTON, M
PHILLIPS, WC
LI, YL
KALATA, K
机构
[1] Brandeis Univ, Waltham
关键词
D O I
10.1107/S0021889892004035
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Software and hardware methods have been developed to correct images for spatial and intensity distortions produced by optical and electro-optical components in X-ray area detectors. Spatial distortions are divided into two types: gross distortions produced by the inherent properties of the detector components and local distortions formed by irregularities in the components. Intensity distortions are separated into three types: those caused by background nonuniformity; those resulting from pixel-dependent nonuniform intensity response; and those resulting from time-dependent variations in background and incident-beam intensity. From background, flat-field, reference and mask images, 'forward' and 'reverse' interpolation tables are generated to correct for spatial distortions and a lookup table is generated to correct for nonuniform sensitivity. The routines have been used successfully on four different area detectors to correct entire images or to correct intensities of individual Bragg peaks. The spatial-distortion correction is good to within 0.1 pixels and the nonuniformity correction to less-than-or-similar-to 2%.
引用
收藏
页码:549 / 558
页数:10
相关论文
共 17 条
[1]   MINIATURE IONIZATION CHAMBERS [J].
ARNDT, UW ;
STUBBINGS, SJ .
JOURNAL OF APPLIED CRYSTALLOGRAPHY, 1988, 21 (05) :577-577
[2]  
ARNDT UW, 1985, METHOD ENZYMOL, V114, P472
[3]   FORMAT ALTERATIONS IN CCD BASED ELECTROOPTIC X-RAY-DETECTORS [J].
DECKMAN, HW ;
GRUNER, SM .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1986, 246 (1-3) :527-533
[4]   CALIBRATION TESTS AND USE OF A NICOLET XENTRONICS IMAGING PROPORTIONAL CHAMBER MOUNTED ON A CONVENTIONAL SOURCE FOR PROTEIN CRYSTALLOGRAPHY [J].
DEREWENDA, Z ;
HELLIWELL, JR .
JOURNAL OF APPLIED CRYSTALLOGRAPHY, 1989, 22 :123-137
[5]   PROTEIN, DNA, AND VIRUS CRYSTALLOGRAPHY WITH A FOCUSED IMAGING PROPORTIONAL COUNTER [J].
DURBIN, RM ;
BURNS, R ;
MOULAI, J ;
METCALF, P ;
FREYMANN, D ;
BLUM, M ;
ANDERSON, JE ;
HARRISON, SC ;
WILEY, DC .
SCIENCE, 1986, 232 (4754) :1127-1132
[6]   CHARACTERISTICS OF A FLAT MULTIWIRE AREA DETECTOR FOR PROTEIN CRYSTALLOGRAPHY [J].
HAMLIN, R ;
CORK, C ;
HOWARD, A ;
NIELSEN, C ;
VERNON, W ;
MATTHEWS, D ;
XUONG, NH .
JOURNAL OF APPLIED CRYSTALLOGRAPHY, 1981, 14 (APR) :85-93
[7]  
HAMLIN R, 1985, METHOD ENZYMOL, V114, P416
[9]  
KALATA K, 1990, P SOC PHOTO-OPT INS, V1345, P270
[10]  
KALATA K, 1985, METHOD ENZYMOL, V114, P486