COMPARISON OF SIMULTANEOUS VERSUS SEQUENTIAL DEFIBRILLATION PULSING TECHNIQUES USING A NONTHORACOTOMY SYSTEM

被引:6
作者
HSIA, HH
KLEIMAN, RB
FLORES, BT
MARCHLINSKI, FE
机构
[1] HOSP UNIV PENN,DIV CARDIOVASC,CLIN CARDIAC ELECTROPHYSIOL LAB,PHILADELPHIA,PA 19104
[2] PHILADELPHIA HEART INST,PHILADELPHIA,PA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1994年 / 17卷 / 07期
关键词
DEFIBRILLATION; NONTHORACOTOMY LEADS; SIMULTANEOUS PATHWAYS; SEQUENTIAL PATHWAYS;
D O I
10.1111/j.1540-8159.1994.tb01489.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The defibrillation threshold (DFT) using simultaneous (SIML) versus sequential (SEQ) pathways for shock delivery was compared in 16 patients with an implanted cardioverter defibrillator. All patients had three-lead nonthoracotomy systems (NTL) using a left chest subcutaneous patch, a right ventricular endocardial lead, and a lead in the coronary sinus (n = 5) or superior vena cava (n = 11). The DFT were determined 2-44 days (17 +/- 17 days) after implantation. The DFT was defined as the lowest energy shock that resulted in successful defibrillation. The first pathway tested was SIML in 12 and SEQ in 4 patients with output beginning at or above the intraoperative DFT, routinely 18 J. The second pathway was tested beginning 2-4 J above the DFT of the first tested pathway. All shocks were delivered in 2-4 J decrement or increment steps. The SEQ pathway shocks resulted in a significantly lower DFT than SIML pathway shocks (14 +/- 6 vs 18 +/- 6 J; P < 0.01). There was no difference in the time delay after ventricular fibrillation initiation before shock delivery for the successful defibrillation between SIML versus SEQ pathways (7 +/- 2 secs for both pathways). In 7 of 16 patients, defibrillation using SEQ pathway resulted in a > 5 J lowering of DFT, while only one patient had > 5 J lowering of DFT using SIML shocks (P <0.05). These results have important implications for selecting the optimal pathway for implantable cardioverter defibrillator therapy with a multilead NTL system.
引用
收藏
页码:1222 / 1230
页数:9
相关论文
共 50 条
[1]   INTRACELLULAR-RECORDING OF INSITU VENTRICULAR CELLS DURING VENTRICULAR-FIBRILLATION [J].
AKIYAMA, T .
AMERICAN JOURNAL OF PHYSIOLOGY, 1981, 240 (04) :H465-H471
[2]   TEMPORAL STABILITY AND PRECISION OF VENTRICULAR DEFIBRILLATION THRESHOLD DATA [J].
BABBS, CF ;
WHISTLER, SJ ;
YIM, GKW .
AMERICAN JOURNAL OF PHYSIOLOGY, 1978, 235 (05) :H553-H558
[3]  
BABBS CF, 1976, MED INSTRUM, V10, P52
[4]   INTRAOPERATIVE COMPARISON OF SEQUENTIAL-PULSE AND SINGLE-PULSE DEFIBRILLATION IN CANDIDATES FOR AUTOMATIC IMPLANTABLE DEFIBRILLATORS [J].
BARDY, GH ;
STEWART, RB ;
IVEY, TD ;
GRAHAM, EL ;
ADHAR, GC ;
GREENE, HL .
AMERICAN JOURNAL OF CARDIOLOGY, 1987, 60 (07) :618-624
[5]   PROSPECTIVE EVALUATION OF INITIALLY INEFFECTIVE DEFIBRILLATION PULSES ON SUBSEQUENT DEFIBRILLATION SUCCESS DURING VENTRICULAR-FIBRILLATION IN SURVIVORS OF CARDIAC-ARREST [J].
BARDY, GH ;
IVEY, TD ;
JOHNSON, G ;
STEWART, RB ;
GREENE, HL .
AMERICAN JOURNAL OF CARDIOLOGY, 1988, 62 (10) :718-722
[6]   A PROSPECTIVE RANDOMIZED EVALUATION OF BIPHASIC VERSUS MONOPHASIC WAVEFORM PULSES ON DEFIBRILLATION EFFICACY IN HUMANS [J].
BARDY, GH ;
IVEY, TD ;
ALLEN, MD ;
JOHNSON, G ;
MEHRA, R ;
GREENE, HL .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 14 (03) :728-733
[7]   PROSPECTIVE COMPARISON OF SEQUENTIAL PULSE AND SINGLE PULSE DEFIBRILLATION WITH USE OF 2 DIFFERENT CLINICALLY AVAILABLE SYSTEMS [J].
BARDY, GH ;
IVEY, TD ;
ALLEN, MD ;
JOHNSON, G ;
GREENE, HL .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 14 (01) :165-171
[8]   RESULTS AND REALISTIC EXPECTATIONS WITH TRANSVENOUS LEAD SYSTEMS [J].
BLOCK, M ;
HAMMEL, D ;
ISBRUCH, F ;
BORGGREFE, M ;
WIETHOLT, D ;
HACHENBERG, T ;
SCHELD, HH ;
BREITHARDT, G .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1992, 15 (04) :665-670
[9]  
BOURLAND JD, 1986, MED INSTRUM, V20, P138
[10]   RELATIONSHIP OF LEFT-VENTRICULAR MASS TO DEFIBRILLATION THRESHOLD FOR THE IMPLANTABLE DEFIBRILLATOR - A COMBINED CLINICAL AND ANIMAL STUDY [J].
CHAPMAN, PD ;
SAGAR, KB ;
WETHERBEE, JN ;
TROUP, PJ .
AMERICAN HEART JOURNAL, 1987, 114 (02) :274-278