Acute and chronic effects of self-expanding nitinol stents in porcine coronary arteries

被引:28
作者
Hong, MK
Beyar, R
Kornowski, R
Tio, FO
Bramwell, O
Leon, MB
机构
[1] WASHINGTON HOSP CTR,DEPT INTERNAL MED,DIV CARDIOL,WASHINGTON,DC 20010
[2] RAMBAM MED CTR,DIV CARDIOL,HAIFA,ISRAEL
[3] UNIV TEXAS,DEPT PATHOL,SAN ANTONIO,TX 78285
关键词
self-expanding stent; porcine coronary artery; long-term patency; histology;
D O I
10.1097/00019501-199701000-00007
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background There is a paucity of experimental data regarding self-expanding stents, This study evaluated the acute and chronic effects of CardioCoil(TM), a self-expanding nitinol coil stent, in porcine coronary arteries. Methods Twenty-three self-expanding nitinol stents were implanted without associated balloon angioplasty in normal coronary arteries of 12 pigs, which were serially sacrificed up to 6 months, Angiographic and histologic analyses were performed to evaluate the deployment characteristics, patency rates, neointimal response, and unique features of the self-expanding nature of the CardioCoil(TM) stent. Results All stents were successfully deployed and remained patent acutely. Three undersized stents migrated proximally and there was one episode of subacute thrombosis in an oversized stent. The remaining stents were patent throughout the survival period and neointimal responses were favorable for up to 6 months (all mean neointima < 200 mu m up to 6 months), There was evidence of continuing stent expansion over time (stent diameter 2.85 +/- 0.78 mm immediately after deployment and 3.24 +/- 0.97 mm at follow-up) and the majority of stent struts were in the adventitia by 6 months, Re-endothelization occurred starting one week after implantation and was complete by 8 weeks, Conclusions This study shows that the CardioCoil(TM) self-expanding nitinol coil stent, is associated with favorable deployment characteristics and patency rates, although appropriate sizing is more crucial than with balloon-expandable stents, More importantly, there appears to be a 'dissociation' between the deep vessel wall injury by the chronic strut expansion process and the neointimal reaction, unlike balloon-expandable stents.
引用
收藏
页码:45 / 48
页数:4
相关论文
共 14 条
[1]   SELF-EXPANDABLE NITINOL STENT FOR CARDIOVASCULAR APPLICATIONS - CANINE AND HUMAN-EXPERIENCE [J].
BEYAR, R ;
SHOFTI, R ;
GRENEDIER, E ;
HENRY, M ;
GLOBERMAN, O ;
BEYAR, M .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1994, 32 (02) :162-170
[2]   CORONARY-ARTERY STENTING AFTER ANGIOPLASTY WITH SELF-EXPANDING PARALLEL WIRE METALLIC STENTS [J].
BONAN, R ;
BHAT, K ;
LEFEVRE, T ;
LEMARBRE, L ;
PAIEMENT, P ;
WOLFF, R ;
LEUNG, TK .
AMERICAN HEART JOURNAL, 1991, 121 (05) :1522-1530
[3]   NON-SURGICAL PLACEMENT OF ARTERIAL ENDOPROSTHESES - A NEW TECHNIQUE USING NITINOL WIRE [J].
CRAGG, A ;
LUND, G ;
RYSAVY, J ;
CASTANEDA, F ;
CASTANEDAZUNIGA, W ;
AMPLATZ, K .
RADIOLOGY, 1983, 147 (01) :261-263
[4]   A RANDOMIZED COMPARISON OF CORONARY-STENT PLACEMENT AND BALLOON ANGIOPLASTY IN THE TREATMENT OF CORONARY-ARTERY DISEASE [J].
FISCHMAN, DL ;
LEON, MB ;
BAIM, DS ;
SCHATZ, RA ;
SAVAGE, MP ;
PENN, I ;
DETRE, K ;
VELTRI, L ;
RICCI, D ;
NOBUYOSHI, M ;
CLEMAN, M ;
HEUSER, R ;
ALMOND, D ;
TEIRSTEIN, PS ;
FISH, RD ;
COLOMBO, A ;
BRINKER, J ;
MOSES, J ;
SHAKNOVICH, A ;
HIRSHFELD, J ;
BAILEY, S ;
ELLIS, S ;
RAKE, R ;
GOLDBERG, S .
NEW ENGLAND JOURNAL OF MEDICINE, 1994, 331 (08) :496-501
[5]   MULTICENTER INVESTIGATION OF CORONARY STENTING TO TREAT ACUTE OR THREATENED CLOSURE AFTER PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY - CLINICAL AND ANGIOGRAPHIC OUTCOMES [J].
GEORGE, BS ;
VOORHEES, WD ;
ROUBIN, GS ;
FEARNOT, NE ;
PINKERTON, CA ;
RAIZNER, AE ;
KING, SB ;
HOLMES, DR ;
TOPOL, EJ ;
KEREIAKES, DJ ;
HARTZLER, GO .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 22 (01) :135-143
[6]   SELF-EXPANDABLE AND HIGHLY FLEXIBLE NITINOL STENT - IMMEDIATE AND LONG-TERM RESULTS IN DOGS [J].
GRENADIER, E ;
SHOFTI, R ;
BEYAR, M ;
LICHTIG, H ;
MORDECHOWITZ, D ;
GLOBERMAN, O ;
MARKIEWICZ, W ;
BEYER, R .
AMERICAN HEART JOURNAL, 1994, 128 (05) :870-878
[7]  
KASAS SP, 1992, J AM COLL CARDIOL, V20, P467
[8]   SELF-EXPANDING ENDOVASCULAR PROSTHESIS - AN EXPERIMENTAL-STUDY [J].
ROUSSEAU, H ;
PUEL, J ;
JOFFRE, F ;
SIGWART, U ;
DUBOUCHER, C ;
IMBERT, C ;
KNIGHT, C ;
KROPF, L ;
WALLSTEN, H .
RADIOLOGY, 1987, 164 (03) :709-714
[9]   BALLOON-EXPANDABLE INTRACORONARY STENTS IN THE ADULT DOG [J].
SCHATZ, RA ;
PALMAZ, JC ;
TIO, FO ;
GARCIA, F ;
GARCIA, O ;
REUTER, SR .
CIRCULATION, 1987, 76 (02) :450-457
[10]   RESTENOSIS AND THE PROPORTIONAL NEOINTIMAL RESPONSE TO CORONARY-ARTERY INJURY - RESULTS IN A PORCINE MODEL [J].
SCHWARTZ, RS ;
HUBER, KC ;
MURPHY, JG ;
EDWARDS, WD ;
CAMRUD, AR ;
VLIETSTRA, RE ;
HOLMES, DR .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1992, 19 (02) :267-274