Performance of the platelet function analyser PFA-100® in testing abnormalities of primary haemostasis

被引:149
作者
Harrison, P
Robinson, MSC
Mackie, IJ
Joseph, J
McDonald, SJ
Liesner, R
Savidge, GF
Pasi, J
Machin, SJ
机构
[1] UCL, Dept Haematol, Haemostasis Res Unit, London WC1E 6HX, England
[2] Great Ormond St Hosp Sick Children, Dept Haematol, London WC1N 3JH, England
[3] St Thomas Hosp, Haemophilia Ctr, London, England
[4] Royal Free Hosp, Haemophilia Ctr, London NW3 2QG, England
关键词
platelets; bleeding time; PFA-100 (R); platelet disorders; von Willebrand's disease;
D O I
10.1097/00001721-199901000-00004
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The PFA-100(R) device is a new instrument for the in-vitro testing of platelet function. Primary haemostasis is stimulated by recording the closure time taken for platelets to seal a 150 mu m aperture in the centre of a membrane coated with collagen and either epinephrine or ADP, Patients with type 3 von Willebrand's disease (n = 4) all had infinitely prolonged closure times (> 200 s) with both types of cartridge. A patient with afibrinogenemia exhibited only slightly prolonged closure times of 111 and 166 s for the ADP and epinephrine membranes, respectively. Patients with Glanzmann's thrombasthenia (n = 6) and Bernard Soulier syndrome (n = 2) had grossly prolonged closure times (> 200 s) with both types of cartridges. These results confirmed that the PFA-100(R) system was highly dependent on normal von Willebrand factor, glycoprotein Ib and glycoprotein IIb/IIIa levels but not on plasma fibrinogen. Patients with storage pool disease (n = 6) and Hermansky Pudlak syndrome (n = 7) had prolonged closure times with the epinephrine cartridge. There was no evidence of enhanced platelet function in patients with antiphospholipid syndrome, in sickle-cell disease or thalassemia. However, ingestion of aspirin resulted in a near consistent and significant prolongation of the closure time for the epinephrine cartridge but not for the ADP cartridge in both normal subjects and patients. The test offers a reliable, reproducible, rapid and simple means of assessing high-shear platelet function in vitro. (C) 1999 Lippincott Williams & Wilkins.
引用
收藏
页码:25 / 31
页数:7
相关论文
共 11 条
[1]   THE PRIMARY ANTIPHOSPHOLIPID SYNDROME - MAJOR CLINICAL AND SEROLOGICAL FEATURES [J].
ASHERSON, RA ;
KHAMASHTA, MA ;
ORDIROS, J ;
DERKSEN, RHWM ;
MACHIN, SJ ;
BARQUINERO, J ;
OUTT, HH ;
HARRIS, EN ;
VILARDELLTORRES, M ;
HUGHES, GRV .
MEDICINE, 1989, 68 (06) :366-374
[2]  
Carcao MD, 1998, BRIT J HAEMATOL, V101, P70
[3]   Screening for von Willebrand disease with a new analyzer using high shear stress: A study of 60 cases [J].
Fressinaud, E ;
Veyradier, A ;
Truchaud, F ;
Martin, I ;
Boyer-Neumann, C ;
Trossaert, M ;
Meyer, D .
BLOOD, 1998, 91 (04) :1325-1331
[4]   Comparison of four commercial citrate blood collection systems for platelet function analysis by the PFA-100(TM) system [J].
Heilmann, EJ ;
Kundu, SK ;
Sio, R ;
Garcia, C ;
Gomez, R ;
Christie, DJ .
THROMBOSIS RESEARCH, 1997, 87 (01) :159-164
[5]  
KRATZER MAA, 1985, HAEMOSTASIS, V15, P357
[6]  
KUNDU SK, 1995, SEMIN THROMB HEMOST, V21, P106
[7]   Characterization of an in vitro platelet function analyzer, PFA-100(TM) [J].
Kundu, SK ;
Heilmann, EJ ;
Sio, R ;
Garcia, C ;
Ostgaard, RA .
CLINICAL AND APPLIED THROMBOSIS-HEMOSTASIS, 1996, 2 (04) :241-249
[8]  
MAMMEN EF, 1995, SEMIN THROMB HEMOST, V21, P113
[9]  
Mammen EF, 1998, SEMIN THROMB HEMOST, V24, P195, DOI 10.1055/s-2007-995840
[10]   A comparison of the effects of aspirin on bleeding time measured using the Simplate(TM) method and closure time measured using the PFA-100(TM), in healthy volunteers [J].
Marshall, PW ;
Williams, AJ ;
Dixon, RM ;
Growcott, JW ;
Warburton, S ;
Armstrong, J ;
Moores, J .
BRITISH JOURNAL OF CLINICAL PHARMACOLOGY, 1997, 44 (02) :151-155