Arterial spin labeling in combination with a look-locker sampling strategy:: Inflow turbo-sampling EPI-FAIR (ITS-FAIR)

被引:187
作者
Günther, M [1 ]
Bock, M [1 ]
Schad, LR [1 ]
机构
[1] Deutsch Krebsforschungszentrum, Forschungssschwerpunkt Radiol Diagnost & Therapie, D-69120 Heidelberg, Germany
关键词
arterial spin labeling; perfusion; cerebral blood flow; Look-Locker;
D O I
10.1002/mrm.1284
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Arterial spin labeling (ASL) permits quantification of tissue perfusion without the use of MR contrast agents. With standard ASL techniques such as flow-sensitive alternating inversion recovery (FAIR) the signal from arterial blood is measured at a fixed inversion delay after magnetic labeling. As no image information is sampled during this delay, FAIR measurements are inefficient and time-consuming. In this work the FAIR preparation was combined with a Look-Locker acquisition to sample not one but a series of images after each labeling pulse. This new method allows monitoring of the temporal dynamics of blood inflow. To quantify perfusion, a theoretical model for the signal dynamics during the Look-Locker readout was developed and applied. Also, the imaging parameters of the new ITS-FAIR technique were optimized using an expression for the variance of the calculated perfusion. For the given scanner hardware the parameters were: temporal resolution 100 ms, 23 images, flip-angle 25.4 degrees. In a normal volunteer experiment with these parameters an average perfusion value of 48.2 +/- 12.1 ml/100 g/min was measured in the brain. With the ability to obtain ITS-FAIR time series with high temporal resolution arterial transit times in the range of -138 - 1054 ms were measured, where nonphysical negative values were found in voxels containing large vessels. (C) 2001 Wiley-Liss, Inc.
引用
收藏
页码:974 / 984
页数:11
相关论文
共 34 条
[1]  
[Anonymous], CLIN MAGNETIC RESONA
[2]   THE CHOICE OF PARAMETERS IN AN NMR EXPERIMENT - APPLICATION TO THE INVERSION-RECOVERY T1 METHOD [J].
BAIN, AD .
JOURNAL OF MAGNETIC RESONANCE, 1990, 89 (01) :153-160
[3]   A general kinetic model for quantitative perfusion imaging with arterial spin labeling [J].
Buxton, RB ;
Frank, LR ;
Wong, EC ;
Siewert, B ;
Warach, S ;
Edelman, RR .
MAGNETIC RESONANCE IN MEDICINE, 1998, 40 (03) :383-396
[4]  
Calamante F, 1996, NMR BIOMED, V9, P79, DOI 10.1002/(SICI)1099-1492(199604)9:2<79::AID-NBM399>3.0.CO
[5]  
2-4
[6]   STAR-HASTE: Perfusion imaging without magnetic susceptibility artifact [J].
Chen, Q ;
Siewert, B ;
Bly, BM ;
Warach, S ;
Edelman, RR .
MAGNETIC RESONANCE IN MEDICINE, 1997, 38 (03) :404-408
[7]   A COMPARISON OF ONE-SHOT AND RECOVERY METHODS IN T1-IMAGING [J].
CRAWLEY, AP ;
HENKELMAN, RM .
MAGNETIC RESONANCE IN MEDICINE, 1988, 7 (01) :23-34
[8]   PERFUSION IMAGING [J].
DETRE, JA ;
LEIGH, JS ;
WILLIAMS, DS ;
KORETSKY, AP .
MAGNETIC RESONANCE IN MEDICINE, 1992, 23 (01) :37-45
[9]   QUALITATIVE MAPPING OF CEREBRAL BLOOD-FLOW AND FUNCTIONAL LOCALIZATION WITH ECHO-PLANAR MR-IMAGING AND SIGNAL TARGETING WITH ALTERNATING RADIO-FREQUENCY [J].
EDELMAN, RR ;
SIEWERT, B ;
DARBY, DG ;
THANGARAJ, V ;
NOBRE, AC ;
MESULAM, MM ;
WARACH, S .
RADIOLOGY, 1994, 192 (02) :513-520
[10]   EPISTAR MRI: Multislice mapping of cerebral blood flow [J].
Edelman, RR ;
Chen, G .
MAGNETIC RESONANCE IN MEDICINE, 1998, 40 (06) :800-805