Comparison of the Sideris and Amplatzer Septal Occlusion devices

被引:48
作者
Walsh, KP [1 ]
Tofeig, M [1 ]
Kitchiner, DJ [1 ]
Peart, I [1 ]
Arnold, R [1 ]
机构
[1] Alder Hey Childrens Hosp, Liverpool L12 2AP, Merseyside, England
关键词
D O I
10.1016/S0002-9149(98)01059-5
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The results of transcatheter atrial septal defect (ASD) occlusion with 2 different devices (Sideris adjustable buttoned device vs Amplatzer Septal Occluder) were compared in 2 consecutive series of patients. Comparative outcomes were assessed by whether a device was implanted or not, by complications and fluoroscopy time of implantation, and by the incidence of residual shunting on transthoracic echocardiography at follow-up. The patient and defect characteristics were similar in both groups. Twenty-eight of 33 Sideris devices and 37 of 39 Amplatzer devices were implanted, The fluoroscopy time for the Amplatzer implants was 13.4 minutes (range 8 to 41) compared with 23.7 minutes (range 11 to 60.6) for the Sideris implants (p <0.001). The complete occlusion rate for the Amplatzer device was 93% compared with 44% for the Sideris device at 1 year (p <0.001), In conclusion, the Amplatzer device produces higher occlusion rates of ASDs with shorter fluoroscopy times, (C) 1999 by Excerpta Medico, Inc.
引用
收藏
页码:933 / 936
页数:4
相关论文
共 17 条
[1]  
ARABIA FA, 1993, J THORAC CARDIOV SUR, V106, P886
[2]   TRANSCATHETER CLOSURE OF ATRIAL SEPTAL-DEFECTS [J].
BABIC, UU ;
GRUJICIC, S ;
DJURISIC, Z ;
VUCINIC, M .
LANCET, 1990, 336 (8714) :566-567
[3]  
Bjornstad PG, 1997, CARDIOL YOUNG, V7, P277
[4]   EXPERIMENTAL ATRIAL SEPTAL-DEFECT CLOSURE WITH A NEW, TRANSCATHETER, SELF-CENTERING DEVICE [J].
DAS, GS ;
VOSS, G ;
JARVIS, G ;
WYCHE, K ;
GUNTHER, R ;
WILSON, RF .
CIRCULATION, 1993, 88 (04) :1754-1764
[5]   Transcatheter closure of atrial septal defects by the ''buttoned'' device: Complications and need for surgical revision [J].
Gildein, HP ;
Dabritz, S ;
Geibel, A ;
Sarai, K ;
VazquezJimenez, J ;
Hugel, W ;
Mocellin, R .
PEDIATRIC CARDIOLOGY, 1997, 18 (05) :328-331
[6]   Surgical repair of a large residual atrial septal defect after transcatheter closure [J].
Hekmat, K ;
Mehlhorn, U ;
deVivie, ER .
ANNALS OF THORACIC SURGERY, 1997, 63 (05) :1456-1458
[7]   Clinical impact of transcatheter closure of secundum atrial septal detects with the double umbrella device [J].
Justo, RN ;
Nykanen, DG ;
Boutin, C ;
McCrindle, BW ;
Freedom, RM ;
Benson, LN .
AMERICAN JOURNAL OF CARDIOLOGY, 1996, 77 (10) :889-&
[8]   SECUNDUM ATRIAL SEPTAL-DEFECT - NONOPERATIVE CLOSURE DURING CARDIAC-CATHETERIZATION [J].
KING, TD ;
THOMPSON, SL ;
STEINER, C ;
MILLS, NL .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1976, 235 (23) :2506-2509
[9]   TRANSCATHETER CLOSURE OF ATRIAL SEPTAL-DEFECTS - EXPERIMENTAL STUDIES [J].
LOCK, JE ;
ROME, JJ ;
DAVIS, R ;
VANPRAAGH, S ;
PERRY, SB ;
VANPRAAGH, R ;
KEANE, JF .
CIRCULATION, 1989, 79 (05) :1091-1099
[10]   Closure of atrial septal defects by transcatheter devices [J].
Magee, AG ;
Qureshi, SA .
PEDIATRIC CARDIOLOGY, 1997, 18 (05) :326-327