Recordings of eye movements for stimulus control during fMRI by means of electro-oculographic methods

被引:22
作者
Felblinger, J
Muri, RM
Ozdoba, C
Schroth, G
Hess, CW
Boesch, C
机构
[1] UNIV BERN,DEPT MR SPECT & METHODOL,BERN,SWITZERLAND
[2] UNIV BERN,DEPT NEUROL,BERN,SWITZERLAND
[3] UNIV BERN,DEPT NEURORADIOL,BERN,SWITZERLAND
关键词
electro-oculography (EGG); functional MR imaging (fMRI); eye motion; monitoring;
D O I
10.1002/mrm.1910360312
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A method for monitoring eye movements in humans during functional MRI is presented. It is based on the acquisition of electro-oculographic (EOG) signals near one eye. EOG potentials were amplified and converted into an optical signal just outside the head coil, An optical fiber was used for signal transmission from inside the magnet bore to the control room. The EOG sensor was tested during EPI sequences at 1.5 Tesla without contamination of the MR signal. Some flow related artifacts on the EOG were observed inside the magnet, but no additional interactions from the MR sequence. An analysis of the latency, direction, and amplitude of the saccadic eye movements was possible.
引用
收藏
页码:410 / 414
页数:5
相关论文
共 15 条
[1]   BURNS ASSOCIATED WITH PULSE OXIMETRY DURING MAGNETIC-RESONANCE-IMAGING [J].
BASHEIN, G ;
SYROVY, G .
ANESTHESIOLOGY, 1991, 75 (02) :382-383
[2]  
BYFORD GH, 1963, J PHYSL, V168, P14
[3]   RF INTERFERENCE SUPPRESSION IN A CARDIAC SYNCHRONIZATION SYSTEM OPERATING IN A HIGH MAGNETIC-FIELD NMR IMAGING-SYSTEM [J].
DAMJI, AA ;
SNYDER, RE ;
ELLINGER, DC ;
WITKOWSKI, FX ;
ALLEN, PS .
MAGNETIC RESONANCE IMAGING, 1988, 6 (06) :637-640
[4]   SYNCHRONIZATION DEVICE FOR ELECTROCARDIOGRAPHY-GATED ECHO-PLANAR IMAGING [J].
FELBLINGER, J ;
DEBATIN, JF ;
BOESCH, C ;
GRUETTER, R ;
MCKINNON, GC .
RADIOLOGY, 1995, 197 (01) :311-313
[5]   ELECTROCARDIOGRAM ACQUISITION DURING MR EXAMINATIONS FOR PATIENT MONITORING AND SEQUENCE TRIGGERING [J].
FELBLINGER, J ;
LEHMANN, C ;
BOESCH, C .
MAGNETIC RESONANCE IN MEDICINE, 1994, 32 (04) :523-529
[6]  
FELMEE J, 1995, P SMR 3 ANN M NIC, P1226
[7]   RECORDING OF HORIZONTAL SACCADIC EYE-MOVEMENTS - METHODOLOGICAL COMPARISON BETWEEN ELECTROOCULOGRAPHY AND INFRARED REFLECTION OCULOGRAPHY [J].
HESS, CW ;
MURI, R ;
MEIENBERG, O .
NEURO-OPHTHALMOLOGY, 1986, 6 (03) :189-198
[8]   BURNS ASSOCIATED WITH CLINICAL MR EXAMINATIONS [J].
KANAL, E ;
SHELLOCK, FG .
RADIOLOGY, 1990, 175 (02) :585-585
[9]  
KERGER KS, 1992, P SMRM 12 ANN M BERL, P661
[10]  
Maier SE, 1995, P SMR 3 SCI M EXH NI, P497