The effects of motion on the performance of pulse oximeters in volunteers (revised publication)

被引:116
作者
Barker, SJ
Shah, NK
机构
[1] Department of Anesthesiology, University of Arizona, College of Medicine, Atlanta, GA
[2] Department of Anesthesiology, University of California, Irvine, CA
[3] Department of Anesthesiology, University of Arizona, Atlanta, GA
[4] Amer. Society of Anesthesiologists, Atlanta, GA
[5] University of Arizona, College of Medicine, Tucson, AZ 85724-5114
关键词
pulse oximeter; monitoring; oxygen; saturation; motion artifact;
D O I
10.1097/00000542-199701000-00014
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background: Pulse oximetry is considered a standard of care in both the operating room and the postanesthetic care unit, and it is widely used in all critical care settings. Pulse oximeters may fail to provide valid SpO(2) data in various situations that produce low signal-to-noise ratio. Motion artifact is a common cause of oximeter failure and loss of accuracy. This study compares the accuracy and data dropout rates of three current pulse oximeters during standardized motion in healthy volunteers. Methods: Ten healthy volunteers were monitored by three different pulse oximeters: Nellcor N-200, Nellcor N-3000, and Masimo SET (prototype), Sensors were placed on digits 2, 3, and 4 of the test hand, which was strapped to a mechanical motion table, The opposite hand was used as a stationary control and was monitored with the same pulse oximeters and an arterial cannula. Arterial oxygen saturation was varied from 100% to 75% by changing the inspired oxygen concentration. While SpO(2) was both constant and changing, the oximeter sensors were connected before and during motion. Oximeter errors and dropout rates were digitally recorded continuously during each experiment. Results: If the oximeter was functioning before motion began, the following are the percentages of time when the instrument displayed an SpO(2) value within 7% of control: N-200 = 76%, N-3000 = 87%, and Masimo = 99%, When the oximeter sensor was connected after the beginning of motion, the values were N-200 = 68%, N-3000 = 47%, and Masimo = 97%, If the alarm threshold was chosen SpO, less than 90%, then the positive predictive values (true alarms/total alarms) are N-200 = 73%, N-3000 = 81%, and Masimo = 100%. In general, N-200 had the greatest SpO(2) errors and N-3000 had the highest dropout rates. Conclusions: The mechanical motions used in this study significantly affected oximeter function, particularly when the sensors were connected during motion, which requires signal acquisition during motion. The error and dropout rate performance of the Masimo was superior to that of the other two instruments during all test conditions. Masimo uses a new paradigm for oximeter signal processing, which appears to represent a significant advance in low signal-to-noise performance.
引用
收藏
页码:101 / 108
页数:8
相关论文
共 10 条
[1]   STATISTICS AND ETHICS IN MEDICAL-RESEARCH .5. ANALYZING DATA [J].
ALTMAN, DG .
BRITISH MEDICAL JOURNAL, 1980, 281 (6253) :1473-1475
[2]   EFFECTS OF METHEMOGLOBINEMIA ON PULSE OXIMETRY AND MIXED VENOUS OXIMETRY [J].
BARKER, SJ ;
TREMPER, KK ;
HYATT, J .
ANESTHESIOLOGY, 1989, 70 (01) :112-117
[3]  
ELFADEL IM, 1995, J CLIN MONITOR, V114, P262
[4]   A PROSPECTIVE-STUDY OF INTRAOPERATIVE PULSE OXIMETRY FAILURE [J].
FREUND, PR ;
OVERAND, PT ;
COOPER, J ;
JACOBSON, L ;
BOSSE, S ;
WALKER, B ;
POSNER, KL ;
CHENEY, FW .
JOURNAL OF CLINICAL MONITORING, 1991, 7 (03) :253-258
[5]   EFFECT OF MOTION ARTIFACT ON PULSE OXIMETERS - EVALUATION OF 4 INSTRUMENTS AND FINGER PROBES [J].
LANGTON, JA ;
HANNING, CD .
BRITISH JOURNAL OF ANAESTHESIA, 1990, 65 (04) :564-570
[6]   CRYING WOLF - FALSE ALARMS IN A PEDIATRIC INTENSIVE-CARE UNIT [J].
LAWLESS, ST .
CRITICAL CARE MEDICINE, 1994, 22 (06) :981-985
[7]   RANDOMIZED EVALUATION OF PULSE OXIMETRY IN 20,802 PATIENTS .1. DESIGN, DEMOGRAPHY, PULSE OXIMETRY FAILURE RATE, AND OVERALL COMPLICATION RATE [J].
MOLLER, JT ;
PEDERSEN, T ;
RASMUSSEN, LS ;
JENSEN, PF ;
PEDERSEN, BD ;
RAVLO, O ;
RASMUSSEN, NH ;
ESPERSEN, K ;
JOHANNESSEN, NW ;
COOPER, JB ;
GRAVENSTEIN, JS ;
CHRAEMMERJORGENSEN, B ;
WIBERGJORGENSEN, F ;
DJERNES, M ;
HESLET, L ;
JOHANSEN, SH .
ANESTHESIOLOGY, 1993, 78 (03) :436-444
[8]   Predictors of pulse oximetry data failure [J].
Reich, DL ;
Timcenko, A ;
Bodian, CA ;
Kraidin, J ;
Hofman, J ;
DePerio, M ;
Konstadt, SN ;
Kurki, T ;
Eisenkraft, JB .
ANESTHESIOLOGY, 1996, 84 (04) :859-864
[9]  
SEVERINGHAUS JW, 1992, ANESTHESIOLOGY, V76, P1018
[10]  
WILSON S, 1990, Pediatric Dentistry, V12, P228