Evaluation of the spatial accuracy of magnetic resonance imaging-based stereotactic target localization for gamma knife radiosurgery of functional disorders

被引:46
作者
Bednarz, G
Downes, B
Corn, BW
Curran, WJ
Goldman, HW
机构
[1] Thomas Jefferson Univ, Kimmel Canc Ctr, Dept Radiat Oncol, Jefferson Med Coll, Philadelphia, PA 19107 USA
[2] Thomas Jefferson Univ, Kimmel Canc Ctr, Dept Neurosurg, Jefferson Med Coll, Philadelphia, PA 19107 USA
[3] Neurosensory Inst Wills Eye Hosp, Philadelphia, PA USA
关键词
functional disorders; magnetic resonance imaging; stereotactic radiosurgery;
D O I
10.1097/00006123-199911000-00028
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
PURPOSE: This study was undertaken to determine the impact of geometric distortions on the spatial accuracy of magnetic resonance imaging (MRI)-guided stereotactic localization for gamma knife functional radiosurgery. METHOD: The spatial accuracy of MRI was evaluated by comparing stereotactic coordinates of intracranial targets, external fiducials, and anatomic structures defined by computed tomographic and MRI studies of the Radionics skull phantom (Radionics, Inc., Burlington, MA), the Rando head phantom, and II patients who underwent gamma knife functional radiosurgery. The distortion in MRI:was assessed from computed tomographic and MRI fusion studies for these patients, as well as from MRI studies acquired by swapping the direction of the magnetic field gradients for five patients who underwent gamma knife radiosurgery and three patients who underwent MRI-guided frameless surgery. A follow-up program to compare the location of the created lesion with the intended target complemented the analysis. RESULTS: The average difference between computed tomographic and MRI stereotactic coordinates of external fiducials, intracranial targets, and anatomic landmarks was of the order of 1 pixel size (0.9 x 0.9 x 1 mm(3)) along the x, y, and z axes. The average linear scaling along these axes as determined by fusion studies was approximately 0.8% and consistent with a single pixel. The follow-up studies, available for seven patients, revealed good agreement between the location of the created lesion and the intended target. CONCLUSION: The spatial accuracy of an MRI-based localization system can be comparable to computed tomography-based localization with the added benefit of MRI resolution. Both machine- and object-related MRI distortions can be reduced to an acceptable level with contemporary scanners, optimized scanning sequences, and distortion-resistant stereotactic instruments.
引用
收藏
页码:1156 / 1161
页数:6
相关论文
共 19 条
[1]   ANALYSIS OF MACHINE-DEPENDENT AND OBJECT-INDUCED GEOMETRIC DISTORTION IN 2DFT MR IMAGING [J].
BAKKER, CJG ;
MOERLAND, MA ;
BHAGWANDIEN, R ;
BEERSMA, R .
MAGNETIC RESONANCE IMAGING, 1992, 10 (04) :597-608
[2]   MRI distortion and stereotactic neurosurgery using the Cosman-Roberts-Wells and Leksell frames [J].
Burchiel, KJ ;
Nguyen, TT ;
Coombs, BD ;
Szumoski, J .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 1996, 66 (1-3) :123-136
[3]   EFFECTS OF COREGISTRATION OF MR TO CT IMAGES ON MR STEREOTAXIC ACCURACY [J].
COHEN, DS ;
LUSTGARTEN, JH ;
MILLER, E ;
KHANDJI, AG ;
GOODMAN, RR .
JOURNAL OF NEUROSURGERY, 1995, 82 (05) :772-779
[4]   Stereotactic radiosurgery for trigeminal neuralgia: A multiinstitutional study using the gamma unit [J].
Kondziolka, D ;
Lunsford, LD ;
Flickinger, JC ;
Young, RF ;
Vermeulen, S ;
Duma, CM ;
Jacques, DB ;
Rand, RW ;
Regis, J ;
Peragut, JC ;
Manera, L ;
Epstein, MH ;
Lindquist, C .
JOURNAL OF NEUROSURGERY, 1996, 84 (06) :940-945
[5]   A COMPARISON BETWEEN MAGNETIC-RESONANCE-IMAGING AND COMPUTED-TOMOGRAPHY FOR STEREOTAXIC COORDINATE DETERMINATION [J].
KONDZIOLKA, D ;
DEMPSEY, PK ;
LUNSFORD, D ;
HEILBRUN, MP ;
KESTLE, JRW ;
DOLAN, EJ ;
KANAL, E ;
TASKER, RR .
NEUROSURGERY, 1992, 30 (03) :402-407
[6]   IMAGE FUSION FOR STEREOTAXIC RADIOTHERAPY AND RADIOSURGERY TREATMENT PLANNING [J].
KOOY, HM ;
VANHERK, M ;
BARNES, PD ;
ALEXANDER, E ;
DUNBAR, SF ;
TARBELL, NJ ;
MULKERN, RV ;
HOLUPKA, EJ ;
LOEFFLER, JS .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 28 (05) :1229-1234
[7]  
Ludeke K M, 1985, Magn Reson Imaging, V3, P329, DOI 10.1016/0730-725X(85)90397-2
[8]   A universal method for geometric correction of magnetic resonance images for stereotactic neurosurgery [J].
Maciunas, RJ ;
Fitzpatrick, JM ;
Gadamsetty, S ;
Maurer, CR .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 1996, 66 (1-3) :137-140
[9]   Effect of geometrical distortion correction in MR on image registration accuracy [J].
Maurer, CR ;
Aboutanos, GB ;
Dawant, BM ;
Gadamsetty, S ;
Margolin, RA ;
Maciunas, RJ ;
Fitzpatrick, JM .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1996, 20 (04) :666-679
[10]   ON THE PROBLEM OF GEOMETRIC DISTORTION IN MAGNETIC-RESONANCE IMAGES FOR STEREOTAXIC NEUROSURGERY [J].
MICHIELS, J ;
BOSMANS, H ;
PELGRIMS, P ;
VANDERMEULEN, D ;
GYBELS, J ;
MARCHAL, G ;
SUETENS, P .
MAGNETIC RESONANCE IMAGING, 1994, 12 (05) :749-765