Converting dose-length product to effective dose at CT

被引:406
作者
Huda, Walter [1 ]
Ogden, Kent M. [2 ]
Khorasani, Mohammad R. [2 ]
机构
[1] Med Univ S Carolina, Dept Radiol, Charleston, SC 29425 USA
[2] SUNY Upstate Med Univ, Dept Radiol, Syracuse, NY USA
基金
美国国家卫生研究院;
关键词
D O I
10.1148/radiol.2483071964
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To determine effective dose (ED) per unit dose-length product (DLP) conversion factors for computed tomographic (CT) dosimetry. Materials and Methods: A CT dosimetry spreadsheet was used to compute patient ED values and corresponding DLP values. The ratio of ED to DLP was determined with 16-section CT scanners from four vendors, as well as with five models from one manufacturer that spanned more than 25 years. ED-to-DLP ratios were determined for 2-cm scan lengths along the patient axis, as well as for typical scan lengths encountered at head and body CT examinations. The dependence of the ratio of ED to DLP on x-ray tube voltage (in kilovolts) was investigated, and the values obtained with the spreadsheet were compared with those obtained by using two other commercially available CT dosimetry software packages. Results: For 2-cm scan lengths, changes in the scan region resulted in differences to ED of a factor of 30, but much lower variation was obtained for typical scan lengths at clinical head and body imaging. Inter-and intramanufacturer differences for ED/DLP were generally small. Representative values of ED/DLP at 120 kV were 2.2 mu v/mGy (.) cm (head scans), 5.4 mu Sv/mGy (.) cm (cervical spine scans), and 18 mu Sv/mGy (.) cm (body scans). For head scans, ED/DLP was approximately independent of x-ray tube voltage, but for body scans, the increase from 80 to 140 kV increased the ratio of ED to DLP by approximately 25%. Agreement in ED/DLP data for all three software packages was generally very good, except for cervical spine examinations where one software package determined an ED/DLP ratio that was approximately double that of the other two. Conclusion: This article describes a method of providing CT users with a practical and reliable estimate of adult patient EDs by using the DLP displayed on the CT console at the end of any given examination. (C) RSNA, 2008.
引用
收藏
页码:995 / 1003
页数:9
相关论文
共 32 条
[1]  
[Anonymous], 2001, ESSENTIAL PHYS MED I
[2]  
[Anonymous], 2000, UNSCEAR 2000 SOURC E
[3]  
Bongartz G.G.S., 1999, EUROPEAN GUIDELINES, V1999, P16262
[4]   Cancer risks attributable to low doses of ionizing radiation: Assessing what we really know [J].
Brenner, DJ ;
Doll, R ;
Goodhead, DT ;
Hall, EJ ;
Land, CE ;
Little, JB ;
Lubin, JH ;
Preston, DL ;
Preston, RJ ;
Puskin, JS ;
Ron, E ;
Sachs, RK ;
Samet, JM ;
Setlow, RB ;
Zaider, M .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2003, 100 (24) :13761-13766
[5]   Assessment of a theoretical formalism for dose estimation in CT: an anthropomorphic phantom study [J].
Brix, G ;
Lechel, U ;
Veit, R ;
Truckenbrodt, R ;
Stamm, G ;
Coppenrath, EM ;
Griebel, J ;
Nagel, HD .
EUROPEAN RADIOLOGY, 2004, 14 (07) :1275-1284
[6]  
*COMM BIOL EFF ION, 2005, 7 BEIR COMM BIOL EFF
[7]   Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography [J].
Einstein, Andrew J. ;
Henzlova, Milena J. ;
Rajagopalan, Sanjay .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2007, 298 (03) :317-323
[8]  
Galanski M, 2000, RAD EXPOSURE COMPUTE
[9]  
Hart D., 1994, R262 NRPB
[10]   An approach for the estimation of effective radiation dose at CT in pediatric patients [J].
Huda, W ;
Atherton, JV ;
Ware, DE ;
Cumming, WA .
RADIOLOGY, 1997, 203 (02) :417-422