PROGRESS IN THE DEVELOPMENT OF A FLUORESCENT INTRAVASCULAR BLOOD-GAS SYSTEM IN MAN

被引:63
作者
MAHUTTE, CK
SASSOON, CSH
MURO, JR
HANSMANN, DR
MAXWELL, TP
MILLER, WW
YAFUSO, M
机构
[1] Department of Medicine, Long Beach, Veterans Administration Medical Center, Long Beach, CA
[2] University of California, Irvine, CA
[3] Cardiovascular Devices, Inc, Irvine, CA
来源
JOURNAL OF CLINICAL MONITORING | 1990年 / 6卷 / 02期
关键词
Blood gas analysis; Equipment: fiberoptic sensors; Monitoring;
D O I
10.1007/BF02828293
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
In vitro and in vivo animal studies have shown accurate measurements of arterial blood pH (pHa), carbon dioxide tension (PaCO2), and oxygen tension (PaO2) with small intravascular fluorescent probes. Initial human clinical studies showed unexplained intermittent large drops in sensor oxygen tension (PiO2). Normal volunteers were studied to elucidate this problem. In the first part of this study, the probe and cannula were manipulated and the probe configuration and its position within the cannula were varied. The decreases in PiO2 were judged to be primarily due to the sensor touching the arterial wall. Retraction of the sensor tip within the cannula eliminated the problem. In the second part of this study, the accuracy of the retracted probe was evaluated in 4 subjects who breathed varying fractions of inspired oxygen and carbon dioxide. The arterial ranges achieved were 7.20 to 7.59 for pH, 22 to 70 mm Hg for PaCO2, and 46 to 633 mm Hg for PaO2. Linear regression of 48 paired sensor (i) versus arterial values showed pHi = 0.896 pHa + 0.773 (r = 0.98, SEE = 0.017); PiCO2 = 1.05 PaCO2-1.33 (r = 0.98, SEE = 2.4 mm Hg); and PiO2 = 1.09 PaO2-20.6 (r = 0.99, SEE = 21.2 mm Hg). Bias (defined as the mean differences between sensor and arterial values) and precision (SD of differences) were, respectively, -0.003 and 0.02 tor pHi, 0.77 and 2.44 mm Hg for PiCO2, and -2.9 and 25.4 mm Hg for PiO2. The mean in vivo 90% response times for step changes in inspired gas were 2.64, 3.88, and 2.60 minutes, respectively, for pHi, PiCO2, and PiO2. © 1990 Little, Brown and Company.
引用
收藏
页码:147 / 157
页数:11
相关论文
共 19 条
[1]  
Lubbers D.W., Opitz N., Die PCO2-PO2 Opcode: cine neue PCO2-bzw. PO2-Messonde zur Messung des PCO2 oder PO2 von Gasen und Flüssigkeiten, Z Naturforsch [C], 3 U, pp. 532-533, (1975)
[2]  
Peterson J.I., Vurek G.G., Fiber optic sensors tor biomedical applications, Science, 224, pp. 123-127, (1984)
[3]  
Peterson, Fiber optic probe for in vivo measurement of oxygen partial pressure, Anal Chem, 56, pp. 62-67, (1984)
[4]  
Wolfbeis O.S., Furhnger E., Kroneis H., Marsoner H., Fluorimetric analysis. 1. A study on fluorescent indicators tor measuring near neutral (physiological) pH values, Z Anal Chem, 314, pp. 119-124, (1983)
[5]  
Gehrich J.L., Lubbers D.W., Opitz N., Et al., Optical fluorescence and its application to an intravascular blood gas system, IEEE Trans Biomed Eng, 2, pp. 117-132, (1986)
[6]  
Miller W.W., Yafuso M., Yan C.F., Et al., Performance of an in-vivo continuous blood gas monitor with disposable probe, Clin Chem, 33, (1987)
[7]  
Tusa J., Hacker T., Hansmann D.R., Et al., Fiber optic microsensor for continuous in-vivo measurement of blood gases, Optical fibers in medicine II. SPIE, 713, pp. 137-143, (1986)
[8]  
Gardner R.M., Direct blood presbure measurement-dynamic response requirements, Anesthesiology, 54, pp. 227-236, (1981)
[9]  
Altman D.G., Statistics and ethics in medical research. V. Analyzing data, Br Med J, 281, pp. 1473-1475, (1980)
[10]  
Martin W.E., Cheung P.W., Johnson C.C., Wong K.C., Continuous monitoring ot mixed venous oxygen saturation in man, Anesth Analg, 52, pp. 784-793, (1973)