OPTIMAL FLOW PATTERN FOR MECHANICAL VENTILATION OF LUNGS .2. EFFECT OF A SINE VERSUS SQUARE-WAVE FLOW PATTERN WITH AND WITHOUT AN END-INSPIRATORY PAUSE ON PATIENTS

被引:30
作者
DAMMANN, JF
MCASLAN, TC
MAFFEO, CJ
机构
[1] UNIV VIRGINIA,MED CTR,CHARLOTTESVILLE,VA 22903
[2] UNIV MARYLAND,SCH MED,MARYLAND INST EMERGENCY MED,BALTIMORE,MD 21201
关键词
D O I
10.1097/00003246-197809000-00001
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
In four series of patients, the efficiency of ventilation of a sine wave without an end-inspiratory pause was compared to a square wave without a pause, a sine wave with a pause to a square wave with a pause, a sine wave to a sine wave with a pause, and a sine wave with a long pause to one with a short pause. The primary mode of evaluation was through simultaneous airway and arterial argon washout curves. Additional cardiopulmonary measurements were made. Results indicate: (1) a statistically significant improvement in ventilation with a sine wave with a pause; (2) a statistically significant improvement with the longer pause as compared to the short pause.
引用
收藏
页码:293 / 310
页数:18
相关论文
共 23 条
[1]   ANALYSIS OF VENTILATION AND PERFORMANCE ABNORMALITIES BY WASHOUT IN ALVEOLAR AIR AND ARTERIAL BLOOD CONTINUOUS MEASUREMENT OF INERT-GAS [J].
AYRES, SM .
CRITICAL CARE MEDICINE, 1976, 4 (05) :261-264
[2]   EFFECTS OF VARYING RESPIRATORY WAVEFORMS ON GAS EXCHANGE [J].
BERGMAN, NA .
ANESTHESIOLOGY, 1967, 28 (02) :390-&
[3]   UNEVEN VENTILATION OF NORMAL AND DISEASED LUNGS STUDIED BY AN OPEN-CIRCUIT METHOD [J].
BRISCOE, WA ;
COURNAND, A .
JOURNAL OF APPLIED PHYSIOLOGY, 1959, 14 (03) :284-290
[4]   OPTIMAL FLOW PATTERN FOR MECHANICAL VENTILATION OF LUNGS - EVALUATION WITH A MODEL LUNG [J].
DAMMANN, JF ;
MCASLAN, TC .
CRITICAL CARE MEDICINE, 1977, 5 (03) :128-136
[5]   LUNG FUNCTION STUDIES .8. ANALYSIS OF ALVEOLAR VENTILATION BY PULMONARY N-2 CLEARANCE CURVES [J].
FOWLER, WS ;
CORNISH, ER ;
KETY, SS .
JOURNAL OF CLINICAL INVESTIGATION, 1952, 31 (01) :40-50
[6]  
FULHEIHAN S, 1976, ANESTH ANALG, V55, P122
[7]  
HERZOG P, 1968, OPUSC MED, V13, P27
[8]  
INGELSTEDT S, 1972, ACTA ANAESTHESIO S47, P8
[9]  
JANSSON L, 1972, SCAND J RESPIR DIS, V53, P237
[10]   SIMPLE METHOD FOR DETERMINATION OF PERFUSION + VENTILATION-PERFUSION RATIO OF UNDERVENTILATED ELEMENTS ( SLOW COMPARTMENT ) OF LUNG [J].
KLOCKE, RA ;
FARHI, LE .
JOURNAL OF CLINICAL INVESTIGATION, 1964, 43 (12) :2227-&