DEVELOPMENT OF AL-26 ACCELERATOR MASS-SPECTROMETRY FOR BIOLOGICAL AND TOXICOLOGICAL APPLICATIONS

被引:31
作者
BARKER, J
DAY, JP
AITKEN, TW
CHARLESWORTH, TR
CUNNINGHAM, RC
DRUMM, PV
LILLEY, JS
NEWTON, GWA
SMITHSON, MJ
机构
[1] UNIV MANCHESTER,DEPT CHEM,MANCHESTER M13 9PL,LANCS,ENGLAND
[2] SERC LAB,NUCL STRUCT FACIL,WARRINGTON WA4 4AD,ENGLAND
关键词
D O I
10.1016/0168-583X(90)90473-8
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We report the use of a 20 MV tandem Van de Graaff accelerator to conduct Al-26 mass spectrometry with fully stripped aluminium ions, and the development of an ion source preparation for the measurement of Al-26:Al-27 ratios in biological materials. Discrimination of Al-26 from Mg-26 is better than 1:10(9). Al-26 calibration is linear with an estimated detection limit approximately 10(-18) g and limiting Al-26:Al-27 ratio approximately 10(-13). Preliminary biological applications include identification of Al carrier proteins in blood, and investigations of Al uptake by cells in culture and Al speciation in natural waters.
引用
收藏
页码:540 / 543
页数:4
相关论文
共 13 条
[3]   3 LOWEST BOUND-STATES OF MG- [J].
BECK, DR .
PHYSICAL REVIEW A, 1984, 30 (06) :3305-3307
[4]   REFLECTED BEAM SPUTTER SOURCE [J].
BRAND, K .
NUCLEAR INSTRUMENTS & METHODS, 1977, 141 (03) :519-520
[5]  
DAY JP, 1990, MAR IN BIOCH DISC GR
[6]  
GITELMAN HJ, 1989, ALUMINIUM HLTH
[7]   PRODUCTION OF 10-80 KEV NEGATIVE HEAVY-IONS BY CHARGE-EXCHANGE IN NA VAPOR [J].
HEINEMEIER, J ;
HVELPLUND, P .
NUCLEAR INSTRUMENTS & METHODS, 1978, 148 (03) :425-429
[8]  
HEWITT CD, 1987, TRACE ELEMENTS HUMAN, P47
[9]   ATOMIC ELECTRON-AFFINITIES [J].
LINEBERGER, WC .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1976, 23 (02) :934-935
[10]   APPLICATION OF ACCELERATOR MASS-SPECTROMETRY IN ALUMINUM METABOLISM STUDIES [J].
MEIRAV, O ;
SUTTON, RAL ;
FINK, D ;
MIDDLETON, R ;
KLEIN, J ;
WALKER, VR ;
HALABE, A ;
VETTERLI, D ;
JOHNSON, RR .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1990, 52 (3-4) :536-539