DIRECT IMAGING OF MAGNETIC-FIELD GRADIENTS BY GROUP SPIN-ECHO SELECTION

被引:50
作者
POSSE, S [1 ]
机构
[1] UNIV BERN,FAC MED,NMR UNIT,CH-3012 BERN,SWITZERLAND
关键词
D O I
10.1002/mrm.1910250103
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A new image processing method for single‐echo gradient echo imaging is presented which extracts local phase gradient information by k‐space filtering instead of by phased reconstruction and spatial differentiation. It is shown that localphase gradient directions and semiquantitative local phase gradient magnitudes can be directly measured, even in regions where phased image reconstruction suffers from multiple phase foldovers due to strong phase modulations. The directional information thus obtained can be used as a reference to identify and correct phase modulation foldovers i phase maps which may be computed from the same raw data. The method is applied here to measure static magnetic field gradients and illustrates fundamental k‐space signal properties of gradient echo imaging. Based on this concept, image artifacts caused by conventional strong k‐space filtering in gradient echo imaging are discussed. Copyright © 1992 Wiley‐Liss, Inc., A Wiley Company
引用
收藏
页码:12 / 29
页数:18
相关论文
共 25 条
[1]   CORRECTION OF PHASE WRAPPING IN MAGNETIC-RESONANCE IMAGING [J].
AXEL, L ;
MORTON, D .
MEDICAL PHYSICS, 1989, 16 (02) :284-287
[2]   MAGNETIC-FIELD MAPPING AND NMR IMAGING BY MULTI-FREQUENCY SENEX [J].
BRAUN, M ;
JUNG, WI ;
LUTZ, O ;
OESCHEY, R .
ZEITSCHRIFT FUR NATURFORSCHUNG SECTION A-A JOURNAL OF PHYSICAL SCIENCES, 1988, 43 (04) :291-296
[3]   ANALYSIS OF THE MAGNITUDE OF SUSCEPTIBILITY EFFECTS IN DISEASE OF THE BRAIN - IMPLICATIONS FOR MR SPECTROSCOPY [J].
COLLINS, AG ;
BRYANT, DJ ;
YOUNG, IR ;
THOMAS, DGT ;
GILL, SS ;
BYDDER, GM .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1988, 12 (05) :775-777
[4]   SIMPLE PROTON SPECTROSCOPIC IMAGING [J].
DIXON, WT .
RADIOLOGY, 1984, 153 (01) :189-194
[5]   DIRECT FLASH MR IMAGING OF MAGNETIC-FIELD INHOMOGENEITIES BY GRADIENT COMPENSATION [J].
FRAHM, J ;
MERBOLDT, KD ;
HANICKE, W .
MAGNETIC RESONANCE IN MEDICINE, 1988, 6 (04) :474-480
[6]  
GIBNEY GH, 1990, 9TH ANN M SOC MAGN R, P563
[7]   REDUCTION OF T2-STAR DEPHASING IN GRADIENT FIELD ECHO IMAGING [J].
HAACKE, EM ;
TKACH, JA ;
PARRISH, TB .
RADIOLOGY, 1989, 170 (02) :457-462
[8]   THE EFFECTS OF FINITE SAMPLING IN SPIN-ECHO OR FIELD-ECHO MAGNETIC-RESONANCE-IMAGING [J].
HAACKE, EM .
MAGNETIC RESONANCE IN MEDICINE, 1987, 4 (05) :407-421
[9]  
HEDLEY M, 1989, 8TH P ANN M SMRM, P920
[10]  
HEDLY M, 1990, 9TH ANN M SOC MAGN R, P562