REPROGRAMMING OF IMPLANTED PACEMAKER FOLLOWING EXTERNAL DEFIBRILLATION

被引:24
作者
BAROLD, SS
ONG, LS
SCOVIL, J
HEINLE, RA
WRIGHT, T
机构
[1] GENESEE HOSP,DEPT MED,DIV CARDIOL,ROCHESTER,NY 14607
[2] GENESEE HOSP,DEPT BIOMED ENGN,ROCHESTER,NY 14607
[3] UNIV ROCHESTER,SCH MED & DENT,ROCHESTER,NY 14627
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1978年 / 1卷 / 04期
关键词
defibrillation; programmable pacemaker; reprogrumming;
D O I
10.1111/j.1540-8159.1978.tb03514.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
This report describes reprogramming of a pulse generator consequent to cardiac defibrillation. Analysis of the explunted pulse generator revealed normal function. We tested the electrical properties of several defibrillaiors and our studies suggest that erratic electrostatic discharges at the time of paddle application before defibrillation may create a signal sequence capable of reprogrumming the particular pulse generator by activating either the reed switch or internal electronics. Copyright © 1978, Wiley Blackwell. All rights reserved
引用
收藏
页码:514 / 520
页数:7
相关论文
共 9 条
[1]   PHANTOM PACEMAKER PROGRAMMING [J].
FIELDMAN, A ;
DOBROW, RJ .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1978, 1 (02) :166-171
[2]   PACEMAKER FAILURE FOLLOWING EXTERNAL DEFIBRILLATION [J].
GIEDWOYN, JO .
CIRCULATION, 1971, 44 (02) :293-&
[3]   PROTECTION OF IMPLANTED PACEMAKERS FROM EXCESSIVE ELECTRICAL ENERGY OF DC SHOCK [J].
LAU, FYK ;
BILITCH, M ;
WINTROUB, HJ .
AMERICAN JOURNAL OF CARDIOLOGY, 1969, 23 (02) :244-&
[4]  
MEYLER FL, 1973, CARDIAC PACING, P323
[5]   PRELIMINARY EXPERIENCE WITH USE OF A PROGRAMMABLE PACEMAKER [J].
MORSE, D ;
SAMUEL, A ;
FERNANDEZ, J ;
LEMOLE, G ;
PARSONNET, V .
CHEST, 1975, 67 (05) :544-548
[6]  
NEITZERT A, 1975, HERZ KREISLAUF, V7, P455
[7]   PACEMAKER FAILURE FOLLOWING EXTERNAL DEFIBRILLATION [J].
PARSONNET, V .
CIRCULATION, 1972, 45 (05) :1144-+
[8]  
1971, MEDTRONIC NEWS, V2, P3
[9]  
1977, 24 CORD CORP CORD TE